QT designed instrumentation -IVVI rack page- -IsoAmp in/out-



Isolation amplifiers (galvanic isolation 1V/V):
note: these modules have a max bandwidth of dc-100kHz (300kHz optional). Modules S5g and M2g provide dc-200MHz isolation  but dc quality is less.
dual Iso-amp-in module S0 (fits any source slot, default in slot Iso-in)

Used to control sources in an analog way from lock-in function genarator etc.
Signals enter on the BNC and the isolated output voltages are routed (backplane) to the summing module S2
The user can set wirejumpers there to have it control sources of choice.
Typical input level is +/-1V, this will generate an output in the related source-module of 100% of the chosen range (ex.:Voltage Source S3 on 10mV/V wil give 10mV output). Most source modules can go up to 200% of the selected range, clipping of the Iso-amp itself takes place at +/-4Vinput , LEDS indicate clipping at approx +/-2V.
Input impedance is 10Mohm, the BNC is connected to a differential amplifier. Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the input sources and iso-amp itself.
dual Iso-amp-out module M0
(default in slot Iso-out, reading modules in Ma and Mb)
(fits also in slot Sd, reading measurement modules placed in Sb and Sc, optional for 4 measurement channels)

Used to isolate the output signals of measurement modules.
Signals enter on the backplane and the isolated output voltages are present on the BNSs on the front
Typical output level is +/-1V, clipping takes place at +/-4Vinput , LEDS indication starts at approx +/-2V.
Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the input sources and iso-amp itself.
dual Iso-amp-in module(fixed slot)

Here the controlling voltages (from lock-in, DAC etc) enter on the BNC.
Iso1 controls source-slot1, Iso2 controls source-slot 2.
Typical input level is +/-1V, this will generate an output in the related source-module of 100% of the chosen range (ex.:Voltage Source S3 on 10mV/V wil give 10mV output). Most source modules can go up to 200% of the selected range, clipping of the Iso-amp itself takes place at +/-4Vinput , LEDS indicate clipping at approx +/3V.
Input impedance is 10Mohm, the BNC is connected to a differential amplifier. Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the input sources and iso-amp itself.
dual Iso-amp-in module with added x0.01 sum input (fixed slot)

Here the controlling voltages (from lock-in, DAC etc) enter on the BNC.
Iso1 controls source-slot1, Iso2 controls source-slot 2.
Typical input level is +/-1V, this will generate an output in the related source-module of 100% of the chosen range (ex.:Voltage Source S3 on 10mV/V wil give 10mV output). The x0.01 input generates an additional 1%/V on the output of the source-module in source-slot 1. Most source modules can go up to 200% of the selected range, clipping of the Iso-amp itself takes place at +/-4Vinput , LEDS indicate clipping at approx +/3V.
Input impedance is 10Mohm, the BNC is connected to a differential amplifier. Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the input sources and iso-amp itself.
IVVI-DAC-rack: combined 2xIso-amp-in 1xout (fixed slot)

Here the controlling voltages (from lock-in, DAC etc) enter on the BNC. The output signal of the measurement module is presented on the iso-out-BNC.
Iso1 controls source-slot1, Iso2 controls source-slot 2.
Typical input level is +/-1V, this will generate an output in the related source-module of 100% of the chosen range (ex.:Voltage Source S3 on 10mV/V wil give 10mV output). Most source modules can go up to 200% of the selected range, clipping of the Iso-amp itself takes place at +/-4Vinput , LEDS indicate clipping at approx +/3V.
Input impedance is 10Mohm, the BNC is connected to a differential amplifier. Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the input sources and iso-amp itself.
Isolation amplifiers (galvanic isolation 1V/V):
dual Iso-amp-out module(fixed slot)

Here the output voltages from the measure modules enter. Via an analog optical isolation stage they are transferred to the BNC outputs on this module.
Typical output level is +/-1V, to achieve this, the gain in the related measure-module should be set to an appropriate value (ex.:Voltage Measurement M1 to 1000x when measuring +/1mV sweeps). Most measure modules can go up to 300% of the selected range ((=+/-3V). Clipping of the Iso-amp itself takes place at +/-4V output , LEDS indicate clipping at approx +/3V.
Output impedance is 470 ohm. Bandwidth limit switches on the output set a lowpass-filter to suppress out-of-band noise from the measurement and iso-amp itself.