IV2 Control Module Functions: 160495 R.Schouten In order of priority: 1-Control of BiasDAC's (8x) & reading of Polarity switches (8x) The BiasDAC's generate settings for gate voltages, there is a hardware polarity protection switch for each DAC DAC update rate is typical < 5Hz Data for BiasDAC's from input fiber1 Data from switches to output fiber1 2-Overload monitor (on software request) Overloads in the SIGNAL UNIT are latched in the control module latched overloadbits are transmitted & latch is cleared Command from input fiber1 Data to output fiber1 3-Signal Conditioning gain & mode settings for the modules in the SIGNAL UNIT The manual controls on the SIGNAL UNIT have to be in the "remote" setting for this control to be effective. Data from input fiber1 4-Battery check (on software request) measurement of pos. & neg. battery voltage Command from input fiber1 Data to output fiber1 5-Power control Supply switches for DAC/ADC1/ADC2/OPTION (batt.life) (note:SIGNAL UNIT is always powered to protect sample) Data from input fiber1